Blood Gas Analyzer

Product

Overview

A blood gas analyzer measures the chemistry that keeps critical-care decisions moving: how acidic the blood is, how much carbon dioxide and oxygen it carries, and the concentrations of the key electrolytes. It runs on a tiny whole-blood sample drawn directly into a syringe or capillary, returns results in under two minutes, and lives at the bedside or in a stat lab rather than a central chemistry line, because in resuscitation the answer is needed now.

A presented sample sits on the Universal Inlet Port, where the Aspiration Probe draws a metered slug into the flow path. The Peristaltic Pump moves that slug, along with calibrants and wash fluids from the Reagent & Calibration System, across the Sensor Cartridge, a replaceable block of electrodes held at body temperature by the Thermostat Block. For oxygenation parameters that electrodes cannot give, the CO-Oximeter Module reads hemoglobin and its fractions optically, and the Control Electronics assembles everything into a report shown on the Display Panel.

How it works

The measurements are electrochemical. Each parameter has its own electrode in the Sensor Cartridge: a glass Glass pH Electrode develops a voltage that tracks pH, a Severinghaus pCO2 Electrode (Severinghaus) reads CO2 as a pH shift behind a gas-permeable membrane, and a Clark-style pO2 Electrode (Clark) reduces oxygen to produce a current proportional to its partial pressure. The electrolyte channels use ion-selective membranes, all read against a single stable Reference Electrode. Because gas solubility and electrode response both shift with temperature, the cartridge is clamped to 37 °C so results reflect conditions inside the patient.

Accuracy depends on relentless calibration. Between samples the Reagent & Calibration System flushes tonometered buffers of known pH and gas content across the same electrodes, and the controller re-zeros and re-slopes each channel so slow membrane drift never reaches the result. The CO-oximeter works differently: it hemolyzes a sub-sample and measures absorbance across several wavelengths, separating oxygenated, deoxygenated, carboxy- and met-hemoglobin by their distinct spectra. Spent fluid is pushed to waste, the probe is rinsed to prevent carryover, and the analyzer reports its self-check status before accepting the next patient sample.

Bill of materials for Blood Gas Analyzer

10 top-level lines as of r15449
# Item / sub-assembly Part no. Qty/assy Ext. qty Parts Type
1 Sample Inlet Module 4 parts blood-gas-analyzer-sample-inlet 1× 1 0 assembly
2 Peristaltic Pump 3 parts blood-gas-analyzer-peristaltic-pump 1× 1 0 assembly
3 Sensor Cartridge 6 parts blood-gas-analyzer-sensor-cartridge 1× 1 0 assembly
4 Reagent & Calibration System 5 parts blood-gas-analyzer-reagent-system 1× 1 0 assembly
5 Thermostat Block 3 parts blood-gas-analyzer-thermostat 1× 1 0 assembly
6 CO-Oximeter Module 6 parts blood-gas-analyzer-cooximeter 1× 1 0 assembly
7 Control Electronics 7 parts blood-gas-analyzer-control 1× 1 0 assembly
8 Display Panel 4 parts blood-gas-analyzer-display 1× 1 0 assembly
9 Result Printer 3 parts blood-gas-analyzer-printer 1× 1 0 assembly
10 Housing & Enclosure 4 parts blood-gas-analyzer-housing 1× 1 0 assembly

396-word article