Hemodialysis Machine

Product

Overview

A hemodialysis machine removes metabolic waste, excess fluid, and electrolytes from the blood of a patient with failing kidneys. It pumps blood from a vascular access through a hollow-fiber dialyzer where, across a semipermeable membrane, solutes diffuse into a counter-flowing dialysate while a controlled pressure gradient pulls off excess water. Two separate hydraulic circuits, one for blood and one for dialysate, run in parallel through the machine.

Blood handling is the job of the Extracorporeal Blood Circuit. A peristaltic blood pump occludes a disposable tubing segment to push blood through the dialyzer at a set flow, a heparin syringe pump meters anticoagulant, and arterial and venous pressure sensors watch both ends of the loop. An ultrasonic air detector and a solenoid venous clamp form the air-embolism safeguard: if bubbles are sensed, the clamp closes and the pump stops.

How it works

The Dialysate Circuit prepares the dialysate online. Treated water is heated and degassed, then proportioning pumps blend it with acid and bicarbonate concentrates while a conductivity cell verifies the ion concentration against limits. Volumetric balancing chambers match the volume of fresh and spent dialysate exactly, so any deliberate imbalance imposed by the ultrafiltration pump equals the fluid removed from the patient. An optical blood-leak detector on the spent line catches a ruptured dialyzer membrane.

The Control System coordinates the pumps, valves, and protective sensors and enforces the treatment prescription, while the operator sets blood flow, dialysate composition, and fluid-removal targets at the Operator Interface. Everything mounts in the Chassis & Cabinet, a mobile chassis on locking casters. Between treatments the Disinfection System cycle flushes the fluid path with heat and chemical agents to control bacterial and endotoxin load.

Bill of materials for Hemodialysis Machine

7 top-level lines as of r45556
# Item / sub-assembly Part no. Qty/assy Ext. qty Parts Type
1 Chassis & Cabinet 4 parts dialysis-machine-cabinet 1× 1 0 assembly
2 Extracorporeal Blood Circuit 8 parts dialysis-machine-blood-circuit 1× 1 0 assembly
3 Dialysate Circuit 12 parts dialysis-machine-dialysate-circuit 1× 1 0 assembly
4 Control System 4 parts dialysis-machine-control-system 1× 1 0 assembly
5 Operator Interface 2 parts dialysis-machine-ui 1× 1 0 assembly
6 Disinfection System 3 parts dialysis-machine-disinfection 1× 1 0 assembly